TUBA01 The Design of a Precision Mechanical Assembly for a Hard X-ray Polarizer 116
 
  • S.P. Kearney, D. Shu, T.S. Toellner
    ANL, Argonne, Illinois, USA
 
  Funding: Work supported by the U.S. Department of Energy, Office of Science, under Contract No. DE-AC02-06CH11357
Hard x-ray polarisers are commonly applied in synchrotron radiation research to produce photons in a pure polarization state, and as polarization filters to analyse the photon’s polarization state after their interaction with a sample medium. We present the design of a mechanical assembly suitable for a hard X-ray polariser that requires multiple degrees of freedom with the base stage capable of handling at least 2-3 kg loads. The intermediary stages (roll, yaw, and translation directions) consist of commercially available tip/tilt and translational stages (Kohzu Precision Co., LTD). However, the requirements of the pitch stage are much more demanding and require a custom-designed flexure-based rotation stage. The design and analysis of this flexure-based rotation stage will be discussed in this study. This will include FEA analysis of the dynamic response and rotation range capabilities which will then be compared to mechanical performance test results using laser interferometers and accelerometer sensors.
 
slides icon Slides TUBA01 [1.586 MB]  
DOI • reference for this paper ※ https://doi.org/10.18429/JACoW-MEDSI2016-TUBA01  
About • paper received ※ 09 September 2016       paper accepted ※ 20 September 2016       issue date ※ 22 June 2017  
Export • reference for this paper using ※ BibTeX, ※ LaTeX, ※ Text/Word, ※ RIS, ※ EndNote (xml)