Paper | Title | Page |
---|---|---|
WEPH39 | Validation Results for Sirius APU19 Front End Prototype | 290 |
|
||
Funding: Ministry of Science, Technology, Innovation and Communication (MCTIC) A Front End (FE) prototype for a 19-mm period length Adjustable Phase Undulator (APU19) beamline of the new Brazilian 4th-generation synchrotron, Sirius, was assembled in the LNLS metrology building in January 2017 to validate main design concepts. Regarding stability, flow-induced vibration (FIV) investigations were carried out on the water-cooled components, and modal analyses were made on the X-Ray Beam Position Monitor (XBPM) support. As for the vacuum system, final pressure levels were investigated and a vacuum breach was intentionally provoked to verify the performance of the equipment protection system (EPS). In addition, cycling tests of the Photon and Gamma shutters were conducted to verify the FE reliability. Moreover, the three-layer protection system, developed to limit the maximum aperture for the high-power slits, was functionally evaluated. Finally, the results were used to improve the FE to its final design. This paper describes the tests setups and results obtained during the validations. |
||
DOI • | reference for this paper ※ https://doi.org/10.18429/JACoW-MEDSI2018-WEPH39 | |
Export • | reference for this paper using ※ BibTeX, ※ LaTeX, ※ Text/Word, ※ RIS, ※ EndNote (xml) | |