WEPE01 | Combined Fixed Mask, Photon Shutter, Safety Shutter, and Collimator Design for BXDS IVU at the CLS | 309 |
|
||
Funding: Canadian Foundation for Innovation The first shutter assembly outside of the Front End (FE) for Brockhouse X-Ray Diffraction and Scattering Sector (BXDS) beamline required a unique design solution to accommodate all components into required safety shutter position. Located between the IVW high energy wiggler monochromator and POE1 wall, the total envelope size approximated 1m x 0.660m (LxW). Accommodating a smaller space required an alternative shutter design than traditionally used implemented at the CLS. The alternative proposed design combined the collimator (CLM), safety shutter (SSH), photon shutter (PSH) and Fixed Mask (FM) into one chamber. Finite Element Analysis (FEA) was conducted on the FM and PSH assembly to verify that geometric designs were adequate for reasonable operation in the beamline. FEA was used to determine the steady-state thermal and static-structural response in both operating positions. Missteer was analyzed for both operating positions to a maximum of 2.5mm (commonly accepted missteer used at the CLS) from center. Finally, two extreme position (5mm) analyses were completed for determination of potential, but unlikely operating conditions. |
||