TUP31 The Cryogenic Current Comparator at CRYRING@ESR 300
 
  • L. Crescimbeni, D.M. Haider, A. Reiter, M. Schwickert, T. Sieber, T. Stöhlker
    GSI, Darmstadt, Germany
  • H. De Gersem, N. Marsic, W.F.O. Müller
    TEMF, TU Darmstadt, Darmstadt, Germany
  • M. Schmelz, R. Stolz, V. Zakosarenko
    IPHT, Jena, Germany
  • F. Schmidl
    FSU Jena, Jena, Germany
  • T. Stöhlker
    IOQ, Jena, Germany
  • T. Stöhlker, V. Tympel
    HIJ, Jena, Germany
  • V. Zakosarenko
    Supracon AG, Jena, Germany
 
  Funding: Work supported by the BMBF under contract No. 05P21SJRB1.
The Cryogenic Current Comparator (CCC) at the heavy-ion storage ring CRYRING@ESR at GSI provides a calibrated non-destructive measurement of beam current with a resolution of 10 nA or better. With traditional diagnostics in storage rings or transfer lines a non-interceptive absolute intensity measurement of weak ion beams (< 1 µA) is already challenging for bunched beams and virtually impossible for coasting beams. Therefore, at these currents the CCC is the only diagnostics instrumentation that gives reliable values for the beam intensity independently of the measured ion species and without the need for tedious calibration procedures. Herein, after a brief review of the diagnostic setup, an overview of the operation of the CCC with different stored ion beams at CRYRING is presented. The current reading of the CCC is compared to the intensity signal of various standard instrumentations including a Parametric Current Transformer (PCT), an Ionization Profile Monitor (IPM) and the Beam Position Monitors (BPMs). It was shown that the CCC is a reliable instrument to monitor changes of the beam current in the range of nA.
 
DOI • reference for this paper ※ doi:10.18429/JACoW-IBIC2022-TUP31  
About • Received ※ 06 September 2022 — Revised ※ 09 September 2022 — Accepted ※ 11 September 2022 — Issue date ※ 19 November 2022
Cite • reference for this paper using ※ BibTeX, ※ LaTeX, ※ Text/Word, ※ RIS, ※ EndNote (xml)  
 
WEP30 Creation of the First High-Inductance Sensor of the New CCC-Sm Series 469
 
  • V. Tympel, T. Stöhlker
    HIJ, Jena, Germany
  • L. Crescimbeni, D.M. Haider, M. Schwickert, T. Sieber, T. Stöhlker
    GSI, Darmstadt, Germany
  • F. Machalett, T. Stöhlker
    IOQ, Jena, Germany
  • M. Schmelz, T. Schönau, R. Stolz, V. Zakosarenko
    IPHT, Jena, Germany
  • F. Schmidl, T. Schönau, P. Seidel
    FSU Jena, Jena, Germany
  • V. Zakosarenko
    Supracon AG, Jena, Germany
 
  Funding: Supported by the BMBF, project number 05P21SJRB1.
Cryogenic Current Comparators (CCC) are presently used at CERN-AD (100 mm beamline diameter) and in the FAIR project at CRYRING (150 mm beamline diameter) for non-destructive absolute measurement of beam currents below 20 ’A (current resolution 10 nA). Both sensor versions (CERN-Nb-CCC and FAIR-Nb-CCC-XD) use niobium as superconductor for the DC-transformer and magnetic shielding. The integrated flux concentrators have an inductance of below 100 ’H at 4.2 Kelvin. The new Sm-series (smart & small) is designed for a beamline diameter of 63 mm and uses lead for the superconducting shield. The first sensor (IFK-Pb-DCCC-Sm-200) has two core-based pickup coils (2x 100 µH at 4.2 K) and two SQUID units, to eliminate Barkhausen current jumps as part of the low frequency 1/f-noise. During the construction some basic experiments on noise behavior (fluctuation’dissipation theorem, white noise below 2 pA/sqrt(Hz)) and magnetic shielding (flux concentrator and shielding as LC circuit resonance , additional mu-metal shielding) were undertaken, the results of which are presented here. Finally, a current resolution of 500 pA could be achieved in the laboratory.
 
poster icon Poster WEP30 [1.474 MB]  
DOI • reference for this paper ※ doi:10.18429/JACoW-IBIC2022-WEP30  
About • Received ※ 05 September 2022 — Revised ※ 10 September 2022 — Accepted ※ 11 September 2022 — Issue date ※ 12 September 2022
Cite • reference for this paper using ※ BibTeX, ※ LaTeX, ※ Text/Word, ※ RIS, ※ EndNote (xml)