Author: Guerrero, A.
Paper Title Page
WEP23 Assessing the Performance of the New Beam Wire Scanners for the CERN LHC Injectors 443
 
  • S. Di Carlo, W. Andreazza, D. Belohrad, J. Emery, J.C. Esteban Felipe, A. Goldblatt, D. Gudkov, A. Guerrero, S. Jackson, G.O. Lacarrere, M. Martin Nieto, A.T. Rinaldi, F. Roncarolo, C. Schillinger, R. Veness
    CERN, Meyrin, Switzerland
 
  The ability of reliably measuring the transverse beam profile in its injectors is essential for the operation of the LHC. This report aims to assess the reliability, stability, and reproducibility of the new generation of beam wire scanners developed at CERN in the framework of the LHC Injectors Upgrade (LIU). The study includes data from the over 60000 scans performed in 2021 and 2022, with a special focus on reproducibility, investigation of optimal operational settings to ensure a large dynamic range, and evaluation of absolute accuracy through comparison with other instruments present in the injectors.  
poster icon Poster WEP23 [1.590 MB]  
DOI • reference for this paper ※ doi:10.18429/JACoW-IBIC2022-WEP23  
About • Received ※ 06 September 2022 — Revised ※ 10 September 2022 — Accepted ※ 11 September 2022 — Issue date ※ 10 December 2022