MOP33 Beam Current Measurements at the Nano-Ampere Level Using a Current Transformer 121
 
  • M. Xiao
    UMCG, Groningen, The Netherlands
  • S. Brandenburg, M.J. Goethem
    PARTREC, Groningen, The Netherlands
  • T. Delaviere, L. Dupuy, F. Stulle
    BERGOZ Instrumentation, Saint Genis Pouilly, France
 
  In conventional proton therapy (PT) typical beam currents are of the order of 1 nA. At these currents dose monitoring is reliably achieved with an ionization chamber. However, at the very high dose rates used in FLASH irradiations (employing beam currents >100 nA) ionization chambers will exhibit large intensity dependent recombination effects and cannot be used. A possible solution is a current transformer. Here we report on the performance of the LC-CWCT (Bergoz Instrumentation, France) which has been developed to push noise floor of such non-destructive current measurement systems into the nano-ampere range. We present first beam current measurements at the PARTREC cyclotron (Netherlands). Beam currents measured by the LC-CWCT and a Faraday Cup were shown to linearly correlate up to the maximum intensity of 400 nA used in the measurements. For pulsed beams, charge measured by the LC-CWCT linearly correlated with pulse length over the measurement range from 50 to 1000 µs. Measurement noise as low as 2.8 nA was achieved. The results confirm that the LC-CWCT has the potential to be applied in FLASH PT for accurate determination of beam current and macro pulse charge.  
DOI • reference for this paper ※ doi:10.18429/JACoW-IBIC2022-MOP33  
About • Received ※ 05 September 2022 — Revised ※ 09 September 2022 — Accepted ※ 13 September 2022 — Issue date ※ 14 September 2022
Cite • reference for this paper using ※ BibTeX, ※ LaTeX, ※ Text/Word, ※ RIS, ※ EndNote (xml)