TY - CONF AU - Li, J.Q. AU - Chen, Q.S. AU - Fan, K. AU - Tang, K. AU - Tian, P. ED - Schaa, Volker RW ED - Jansson, Andreas ED - Shea, Thomas ED - Olander, Johan TI - Design and Simulation of a Cavity BPM for HUST Proton Therapy Facility J2 - Proc. of IBIC2019, Malmö, Sweden, 08-12 September 2019 CY - Malmö, Sweden T2 - International Beam Instrumentation Conferenc T3 - 8 LA - english AB - In proton therapy facility, non-destructive beam diagnostic devices are essential for on-line measurement during the patient treatment. To meet the clinical requirement, the beam current becomes ultra-low of the order of nano-ampere, which is a great challenge to non-destructive beam diagnostics because of the extremely low signal level. Compared with conventional non-destructive beam diagnostic devices, the cavity beam position monitor (BPM) has a high shunt impedance to get enough power levels, so a cavity BPM system is designed for HUST-PTF. It is made up of two resonant cavities called reference cavity and position cavity, respectively. Both cavities are simulated and optimized by CST Microwave Studio and Particle Studio. Finally, the electronics of cavity BPM we plan to use is shown. PB - JACoW Publishing CP - Geneva, Switzerland SP - 530 EP - 533 KW - cavity KW - proton KW - coupling KW - simulation KW - impedance DA - 2019/11 PY - 2019 SN - 2673-5350 SN - 978-3-95450-204-2 DO - doi:10.18429/JACoW-IBIC2019-WEPP010 UR - http://jacow.org/ibic2019/papers/wepp010.pdf ER -