Author: Gornostaev, P.B.
Paper Title Page
MOPG60 Development, Calibration and Application of New-Generation Dissectors With Picosecond Temporal Resolution 205
 
  • O.I. Meshkov, O. Anchugov, G.Y. Kurkin, A.V. Petrozhitskii, D.A. Shvedov, E.I. Zinin
    BINP SB RAS, Novosibirsk, Russia
  • V.L. Dorokhov
    BINP, Novosibirsk, Russia
  • P.B. Gornostaev, M.Ya. Schelev, E.V. Shashkov, A. V. Smirnov, A.I. Zarovskii
    GPI, Moscow, Russia
 
  A dissector is an electron-optical device designed for measurement of periodic light pulses of subnanosecond and picosecond duration. LI-602 dissector developed at BINP SB RAS is widely used for routine measurements of a longitudinal profile of electron and positron beams at BINP electron-positron colliders and other similar installations. LI-602 dissector is a part of many optical diagnostic systems and provides temporal resolution of about 20 ps. Recently a new generation of picosecond dissectors were created on the basis of the PIF-01/S1 picosecond streak-image tube designed and manufactured at the GPI Photoelectronics Department. The results of the measurements of instrument function of the new dissector based on PIF-01/S1, which were carried out in the static mode, showed that temporal resolution of the dissector can be better than 3-4 ps (FWHM). The results of temporal resolution calibration of the new-generation picosecond dissector, carried out at the specialized set-up based on a femtosecond Ti:sapphire laser, and recent results of longitudinal beam profile measurements at BINP accelerators are given in this work.  
DOI • reference for this paper ※ DOI:10.18429/JACoW-IBIC2016-MOPG60  
Export • reference for this paper using ※ BibTeX, ※ LaTeX, ※ Text/Word, ※ RIS, ※ EndNote (xml)