| Paper | 
        Title | 
        Page | 
    
    
        | PS05 | 
        Recent Improvements of a Cryogenic Current Comparator for nA Ion Beams with High Intensity Dynamics
         | 
        109 | 
    
    
        |   | 
        
                - A. Peters, H. Reeg, P. Forck
 
                       GSI, Gesellschaft für Schwerionenforschung, Darmstadt, Germany 
                - W. Vodel, R. Neubert
 
                       FSU Jena, Institut für Festkörperphysik, Friedrich Schiller Universität, Jena, Germany 
          | 
          | 
    
        
            |   | 
            
            Former measurements of extracted ion beams from the
            GSI heavy ion synchrotron SIS showed large current fluctuations
            in the microsecond region with a high peak-to-average
            ratio. To adapt our Cryogenic Current Comparator
            (CCC) to this time structure the detectors electronics have
            been carefully modified.
            
            The most important improvement of the new DC
            SQUID system affects the enlargement of the bandwidth
            and the slew rate of the measuring system. In addition the
            existing data acquisition system for e.g. SEMs (Secondary
            Emission Monitors) was extended to digitize the CCC signals
            simultaneously. Measurements of Argon beams will
            be shown to demonstrate the improved capabilities of the
            upgraded Cryogenic Current Comparator.
             | 
              | 
        
    
        | PT07 | 
        Emittance measurements at the new UNILAC-pre-stripper using a pepper-pot with a PC-controlled CCD-camera
         | 
        171 | 
    
    
        |   | 
        
                - M. Dolinska, M. Domke, P. Forck, T. Hoffmann, D. Liakin, A. Peters, P. Strehl
 
                       GSI, Gesellschaft für Schwerionenforschung, Darmstadt, Germany 
          | 
          | 
    
        
            |   | 
            
            The complex mathematical algorithms and procedures
            to extract emittance data from intensity distributions
            measured with a single shot pepper-pot device are
            described. First results of mathematical evaluation from
            the commissioning of the new GSI pre-stripper linac
            structures are presented.
             | 
              |