MO6RFP022 | The Design and Test of Plug-In Cryopumps | 399 |
|
||
The design and cryo-test system of a plug-in cryopump used in CYCIAE-100 is introduced. The plug-in cryopump consists of two cryopanels, a baffle, a half-opened shield, and two GM refrigerators (CGR411, CVI) which power is 83W/80K at the first stage and 7.5W/20K at the second stage, its designed pumping speed is 15000L/s. Cryo-test system of plug-in cryopump employs the flux method to test pumping speed, cool-down time, ultimate pressure, temperature distribution on cryopanel and capacity at the pressure of 10-6Pa to 1Pa. The heat load calculation of cryopanel and shield including baffle is conducted in succession. In the end a comparison between design parameters and test results is drawn. |