MOPD086 Beam Position Monitoring Based on Higher Beam Harmonics for Application in Compact Medical and Industrial Linear Electron Accelerators 900
 
  • M. Ruf, L. Schmidt
    U. Erlangen-Nurnberg LHFT, Erlangen
  • S. Setzer
    Siemens Med, Erlangen
 
 

The usability of conventional BPM topologies in compact linear accelerators used for medical and industrial applications is very limited due to tight space restrictions in such systems. To overcome these limitations, a different approach is introduced which is based on integrating the pickups into low-field regions of the accelerating structure and evaluating higher beam harmonics. Applications based on this approach will require RF frontends in frequency ranges beyond those covered by BPM dedicated hardware which is currently commercially available. Therefore, a demonstrator setup is presented which is capable of investigating suitable RF frontends for the proposed method. The demonstrator uses capacitive pickups of the button type for displacement sensing and allows for control of the beam position with the help of feedback steering coils which are typically used for compact linacs. Representative sensitivity measurement results based on the evaluation of the 2nd S-Band beam harmonic are also presented in this paper.