A   B   C   D   E   F   G   H   I   J   K   L   M   N   O   P   Q   R   S   T   U   V   W   X   Y   Z  

Krishnan, R.

Paper Title Page
MOPEA050 Development of a prototype 15 MeV Electron Linac 187
 
  • T.S. Dixit, S.T. Chavan, R. Krishnan, C.S. Nainwad, S.N. Pethe, K.A. Thakur, T. Tiwari, M.M. Vidwans
    SAMEER, Mumbai
  • A. Deshpande
    Sokendai, Ibaraki
 
 

A successful development of a 6 MeV electron radiotherapy machine at SAMEER, India was reported earlier*. Now a 15 MeV electron linac prototype is designed, developed and tested at our site. We have measured a beam current of 80 mA at the X-ray target attached to the linac. Energy gained by electrons in a cavity chain of about 1.2 m length is measured to be more than 15 MeV using a 6 MW klystron power source. An RF window capable of handling 12kW average power is attached to the linac tube and it is cooled by water. The final linac parameters measured were at par with the designed values. A high voltage modulator and control console for the linac are designed and developed in house. This paper will describe key aspects of the design and development process of the complete system. Also future applications are planned like-dual energy dual mode linac for radiotherapy, cargo scanning system and compact compton X-ray source using this technology is briefed in this paper.


* R.Krishnan et. al. "S band linac tube developmental work in SAMEER", FR5REP083, PAC09, Vancouver, Canada.