TUPEA041 Drift Calibration Techniques for Future FELs 1419
 
  • F. Ludwig, C. Gerth, K.E. Hacker, M. Hoffmann, G. Moeller, P. Morozov, Ch. Schmidt
    DESY, Hamburg
  • W. Jalmuzna
    TUL-DMCS, Łódź
 
 

Future FELs (Free-Electron-Lasers) requires a precise detection of the cavity field in the injector section with a resolution of much less than 0.01 deg in phase and 0.01% in amplitude for a cavity operation frequency at 1.3GHz. Long-term stable SASE (Self Amplified Spontaneous Emission) operation mainly suffers from injector accelerator components and the stability of the reference distribution. Especially thermal instabilities of the distributed cavity field detectors, probe pickup cables and their mechanical vibrations influence the energy stability dramatically on a scale of 0.1%, a scale which is 10 times worse than required. To eliminate the long-term amplitude and phase changes, we injected a reference signal prior to the arrival of the cavity field signal. This enabled pulse-to-pulse calibration which compensated for the drifts of the field detectors. We demonstrated a dramatic phase and amplitude stability improvement from the ps-range to the 0.008 deg (peak-to-peak) range in phase and 0.02% (peak-to-peak) in amplitude; this represents an improvement in drifts by a factor of about 100. The injected calibration was successfully employed during FLASH operation.