Author: Grossmann, M.
Paper Title Page
TUC3O04 Reusable Patient Safety System Framework for the Proton Therapy Centre at PSI 549
 
  • P. Fernandez Carmona, M. Eichin, M. Grossmann, E. Johansen, A. Mayor, H.A. Regele
    PSI, Villigen PSI, Switzerland
 
  A new gantry for can­cer treat­ment is being in­stalled at the Pro­ton Ther­apy Cen­tre in the Paul Scher­rer In­sti­tut (PSI), where al­ready two gantries and a fixed line op­er­ate. A pro­tec­tion sys­tem is re­quired to en­sure the safety of pa­tients, re­quir­ing stricter re­dun­dancy, ver­i­fi­ca­tion and qual­ity as­sur­ance (QA) mea­sures than other ac­cel­er­a­tors. It su­per­vises the Ther­apy Sys­tem, sen­sors, mon­i­tors and op­er­a­tor in­ter­face and can ac­tu­ate mag­nets and beam block­ers. We built a reusable frame­work to in­crease the main­tain­abil­ity of the sys­tem using the com­mer­cial IFC1210 VME con­troller, de­vel­oped for other PSI fa­cil­i­ties. It fea­tures a FPGA im­ple­ment­ing all the safety logic and two proces­sors, one ded­i­cated to de­bug­ging and the other to in­te­grat­ing in the fa­cil­ity's EPICS en­vi­ron­ment. The frame­work per­mit­ted us to re­duce the de­sign and test time by an es­ti­mated 40% thanks to a mod­u­lar ap­proach. It will also allow a fu­ture ren­o­va­tion of other areas with min­i­mum ef­fort. Ad­di­tion­ally it pro­vides built-in di­ag­nos­tics such as time mea­sure­ment sta­tis­tics, in­ter­lock analy­sis and in­ter­nal vis­i­bil­ity. The au­toma­tion of sev­eral tasks re­duces the bur­den of QA in an en­vi­ron­ment with tight time con­straints.  
slides icon Slides TUC3O04 [10.390 MB]  
Export • reference for this paper using ※ BibTeX, ※ LaTeX, ※ Text/Word, ※ RIS, ※ EndNote (xml)