Author: Sumitomo, H.
Paper Title Page
MOPB52 Status and Activities of the SPring-8 Diagnostics Beamlines 186
 
  • S. Takano, M. Masaki, A. Mochihashi, H. Ohkuma, M. Shoji, K. Tamura
    JASRI/SPring-8, Hyogo-ken, Japan
  • H. Sumitomo, M. Yoshioka
    SES, Hyogo-pref., Japan
 
  At SPring-8 synchrotron radiation (SR) in both the X-ray and the visible bands is exploited in the two diagnostics beamlines. The diagnostics I beamline has a dipole magnet source. The beam size is measured by imaging with the zoneplate X-ray optics. Recently, the transfer line of the visible light has been upgraded. The in-vacuum mirror was replaced to increase the acceptance of the visible photons. A new dark room was built and dedicated to the gated photon counting system for bunch purity monitoring. To improve the performance, the input optics of the visible streak camera was replaced by a reflective optics. Study of the power fluctuation of visible SR pulse is in progress to develop a diagnostic method of short bunch length. The diagnostics II has an insertion device (ID). To monitor stabilities of the ID photon beam, a position monitor for the white X-ray beam based on a CVD diamond screen was installed. A turn-by-turn diagnostics system using the monochromatic X-ray beam was developed to observe fast phenomena such as beam oscillation at injection for top-up and beam blowups caused by instabilities. Study of temporal resolution of the X-ray streak camera is also in progress.