Author: Pernizki, A.
Paper Title Page
MOPB77 Adjustable Optics for a Non-destructive Beam Profile Monitor based on Scintillation of Residual Gas 253
 
  • K. Reimers, V. Kamerdzhiev, A. Pernizki
    FZJ, Jülich, Germany
 
  The scintillation profile monitor (SPM) is being developed at COSY in addition to the existing ionisation profile monitor (IPM). Contrary to the IPM it does not require in-vacuum components, making it a robust and inexpensive instrument. The SPM is suitable for high intensity operation rather than operation with low intensity polarised beams. A multichannel PMT is used to detect scintillation light. The rate of detectable scintillation events is about three orders of magnitude lower compared to the rate of ionisation events. To boost the photon yield, small amounts of nitrogen are injected into the SPM vacuum chamber. An adjustable light focusing system is being built to optimise the SPM performance for different machine operation modes. The new system allows using a variety of optical components ranging from single lenses to high-grade camera objectives. Cylindrical lenses are considered to further boost the sensitivity by better fitting the beam image to the detector geometry. The latest experimental results and the new design of the optical system are presented.