- A. Miura, T. Maruta
JAEA/J-PARC, Tokai-mura, Japan
- K. Hasegawa, N. Ouchi, H. Sako
JAEA, Ibaraki-ken, Japan
- Z. Igarashi, M. Ikegami, T. Miyao
KEK, Ibaraki, Japan
|
|
|
Ar gas proportional BLMs have measured the beam loss through operations, but they are also sensitive to background noise of X-ray emitted from RF cavities. We have tried to measure the beam loss using scintillation monitors which would bring more accurate beam loss measurements with suppression of X-ray noise. We measured beam loss using scintillation beam loss monitors. Because this scintillation BLM is sensitive for low energy gamma-rays and fast neutrons, small signals from X-rays would be also detected. As the measurement results, a good signal to noise ratio is observed for the scintillation monitor with quite low sensitivity to the background X-ray. And many single events are observed in the intermediate pulse bunch with about 600 ns as pulse width. After all signals passed through the filter circuit and integrated circuit, total amount of X-ray noise can become smaller. We obtained the good performances of scintillation BLM with small effect of X-ray noise. This monitor can be used for beam loss measurement and a knob for tuning. Furthermore, because the detail structure can be detected, this monitor could be employed for another diagnostic device.
|
|