Paper | Title | Page |
---|---|---|
TUPD19 | Initial Tests of New Electron and Photon Beam Position Monitor Electronics at the Advanced Photon Source | 344 |
|
||
Funding: Use of the APS, an Office of Science User Facility operated for the U.S. Department of Energy (DOE) Office of Science by ANL, was supported by the U.S. DOE under Contract No. DE-AC02-06CH11357 Measurements were done at the Advanced Photon Source (APS) with Libera Brilliance+, connected to the small-aperture insertion device vacuum chamber pickup electrodes near the beamline 35-ID source point. A photoemission-based photon beam position monitor located 16.35 meters downstream of the center of the ID straight section was also monitored using Libera Photon electronics in horizontal/vertical configuration. Top-up injection transients were recorded simultaneously on both units, providing details about the electron and photon beam motions before, during, and after injection in the storage ring and beamline front end. FFT spectra from the APS-developed BSP-100 broadband BPM data acquisition electronics were compared with the Libera instruments. This article discusses the calibration procedure for electron and photon beam position monitors along with results of these measurements. |
||