TUPD15 Technology Selection for the Beam Position Tuning System in Hadrontherapy Facilities 332
 
  • C. Belver-Aguilar, C. Blanch Gutierrez, A. Faus-Golfe, J.J. García-Garrigós
    IFIC, Valencia, Spain
  • E. Benveniste, M. Haguenauer, P. Poilleux
    LLR, Palaiseau, France
 
  Funding: CYCIT – IN2P3: AIC10-D-000518
The Beam Delivery System of some hadrontherapy facilities is characterized by having scanning magnets, which move the beam in order to irradiate all the tumor volume. To control the beam position, a Beam Position Monitor (BPM) is needed. The BPM described in this paper is a new type of BPM based on four scintillating fibers coupled to four photodiodes to detect the light produced by the fibers when intercepting the beam. We present here the study of the possible photodiodes able to read the light emitted by the scintillating fiber, and the tests performed in order to find the most suitable photodiode to measure the beam position from the variations in the beam current. The setup used for the tests comprises a Sr-90 source, which emits electrons, a scintillating fiber, converting these electrons into photons, and a photodiode, which detects the photons leaving the fiber. The photodiodes studied have been of two types: Avalanche Photodiode (APD) and Multi Pixel Photon Counter (MPPC). In this paper both photodiodes are compared and the results are presented.