TUPD83 Photodiode Calibration using an Electrical Substitution Radiometer in the Hard X-ray Region 500
 
  • N.I. Bolibruch, R. Igarashi, J.M. Vogt
    CLS, Saskatoon, Saskatchewan, Canada
 
  Funding: Work supported by NSERC, NRC, CIHR, WEDC.
An electrical substitution radiometer under development at the Canadian Light Source (CLS) has been used to calibrate a photodiode (AXUV100) from International Radiation Detectors Inc. within an energy range of 8 keV to 30 keV. These measurements were made using monochromatic X-rays on the Biomedical Imaging and Therapy bend magnet beam line and the Hard X-Ray Microanalysis beam line at the CLS. The results were then compared with silicon absorption calculations using data from the NIST mass absorption coefficient tables. Good agreement has been found between the diode calibration obtained from the radiometer and the theoretical calculation of the diode response.