TUPTPF044 Beam Quality Measurements at the Synchrotron and HEBT of the Heidelberg Ion Therapy Center 210
 
  • T. Hoffmann, D. Ondreka, A. Reiter, M. Schwickert
    GSI, Darmstadt
  • A. Peters
    HIT, Heidelberg
 
 

The Heidelberg Ion Therapy Center (HIT) for tumor treatment is presently being commissioned using the beam diagnostic devices designed and produced by the GSI beam diagnostic department. To fulfil the requirements for hadron therapy a high-resolution analysis of the particle distribution within the slowly extracted beam is necessary. We present spill-structure measurements for carbon ion beams at energies from 88 MeV up to 430 MeV, also with respect to the spill-pause and abort functionality of the rf-knock-out extraction method. The spill-structure, as measured by internal intercepting ionization chambers (IC) is compared to data recorded with external beam loss monitors (BLM). The high-resolution data acquisition system with sampling rates up to 10 kSa/s and the connected detectors are described and the achievements during the commissioning phase are discussed.